Page 1 - STDbus Sensor Coprocessor

Sensoray Model 7429STDbus Sensor Coprocessor

Page 3

HARDWARE CONFIGURATIONDEFAULTSThe 7429 requires installation of programming shunts to select various options such as I/O port addresses, interrupts, a

Page 4 - Special Handling Instructions

I/O PORT MAPPINGThe 7429 occupies two consecutive port addresses in the selected STD I/O space. These ports may be mapped to any address from 00H to

Page 5

0 I I I2 I I R4 I R I6 I R R8 R I IA R I RC R R IE R R RI indicates shunt should be insertedR indicates shunt should be removedMSN is the most signifi

Page 6 - INTRODUCTION

ALARM interrupt E19 enables alarm interruptSystem interrupt enable E18 routes interrupt request to STD busHARDWARE FILTERSEach channel is provided wit

Page 7

PROGRAMMINGCOMMUNICATIONProgramming of the 7429 is achieved via a set of built-in commands which the 7429's onboard microcomputer is programmed t

Page 8

The CRMT and DAV bits are used for communication handshake control as described in the next section. During a coprocessor reset, CRMT is held low — t

Page 9

'' BasePort% is globally visible within this module'COMMON SHARED BasePort%'7429 base port address'' This subprogram est

Page 10

COMMANDSThe great versatility of the 7429 is due to its simple but powerful set of built-in com-mands. Commands vary in length depending on the need

Page 11 - HARDWARE CONFIGURATION

Define SensorThis command declares the type of sensor connected to a particular channel. Each sup-ported sensor type is assigned a unique 8-bit senso

Page 12 - I/O PORT MAPPING

TABLE OF CONTENTSBASICS...1WAR

Page 13 - INTERRUPTS

Example: A 350¾ pressure gage is to be connected to channel 7. The gage spec-ification is 3 mV/V at 150 PSI. 3 (mV/V) x 10 = 30.The application req

Page 14 - HARDWARE FILTERS

Read DataThis command returns the most recently measured sensor data from the specified chan-nel. The returned value is scaled according to the decla

Page 15 - PROGRAMMING

Set Alarm LimitsThis command declares upper and lower alarm limits for the specified channel. After declaring limits, an alarm will "sound"

Page 16 - STATUS REGISTER

Read AlarmsThis command returns the status the channel alarms. Alarm flags will be returned from either channels 0-7 or 8-15, depending on the comman

Page 17

Read Board TemperatureThis command returns the temperature of the thermocouple cold-junction compensa-tion sensors (if 7409TB termination boards are c

Page 18 - COMMANDS

Set Open Sensor Data ValuesThis command establishes the data value to be returned for each channel if an open sen-sor is detected. Depending on the f

Page 19

Set Filter Time ConstantThis command establishes a software single-pole low-pass filter for the specified chan-nel. The second command byte specifies

Page 20

Tare GageThis command tares the strain/pressure gage connected to the specified channel. Gage taring is commonly used to subtract off the load offset

Page 21

Activate 50 Hz Rejection FilterThis command will cause the 7429 to reject 50 Hertz differential noise on all channels. Upon power-up reset, the 7429

Page 22

Read All ChannelsThis command returns the most recently acquired data from a block of eight sensor channels. Depending on the command form used, data

Page 23

Activate 50 Hz Rejection Filter... ...20Read All Channels ...

Page 24

Calibrate BoardThis command is used to calibrate all 7429 internal standards. All measurements made by the coprocessor are referenced to these standa

Page 25

NEXT calCode%

Page 26 - RESPONSE: NONE

CONNECTIONSGENERALAll sensors are connected to the 7429 through the 40-pin ribbon cable connectors P1 and P2. Optionally, sensors may be connected to

Page 27

no current passes through the sense lines, therefore no voltage develops across the sense wires.Three-wire connectionUsing three wires has the advanta

Page 28 - COMMAND: (128)

DC voltage connectionsThere is no need for thermocouples to connect to the POWER+ or POWER- pins be-cause thermocouples contain their own built-in pow

Page 29

THERMISTORSThermistors (type 10K) have much higher resistance values than RTD's over most of their operating range, therefore a connection identi

Page 30

4-20 mA current loop connectionsSense+Sense-ShieldSensoray P/N 7408R(+)(-)

Page 31 - NEXT calCode%

CUSTOM RESISTIVE SENSORSThe 7429 possesses the capability of interfacing custom resistive sensors to any sensor channel with the following limitations

Page 32 - CONNECTIONS

D = 1R2 + 17R - 3105where:D is the displacement in millimetersR is the transducer resistance in ohms

Page 33 - THERMOCOUPLES AND DC VOLTAGE

This QuickBasic code segment invokes the define channel sensor command for this re-sistive linear displacement transducer:CALL SendByte (16 + 5)'

Page 34 - Thermocouple connections

Limited WarrantySensoray Company, Incorporated (Sensoray) warrants the model 7429 hardware to be free from defects in material and workmanship and per

Page 35 - STRAIN AND PRESSURE GAGES

THERMOCOUPLE THEORYThermocouples are typically formed from two wires having dissimilar metals. The wires are built into a cable having a heat-resistan

Page 36

COMMAND SUMMARYDEFINE CHANNEL SENSOR - declare sensor type connected to a channelCustom resistive sensors only:command: (16 + CHAN),(SENSOR DEFINITION

Page 37 - CUSTOM RESISTIVE SENSORS

READ BOARD TEMPERATURE - read 7409TB temperature7409TB (channels 0-7):command: (64)response: (TEMP MSB),(TEMP LSB)7409TB (channels 8-15):command: (65)

Page 38 - + 17R - 3105

SET OPEN SENSOR VALUES - establish data value in event of open sensorChannels 0-7:command: (80),(OPEN SENSOR FLAGS)response: NONEChannels 8-15:command

Page 39

SENSOR TABLESSensor definition codes and data scale factors:SENSOR TYPE CODE DATA SCALING per bitVoltage±100 mV f.s. 17H 5 µV±500 mV f.s 16H 20 µV±5

Page 40 - THERMOCOUPLE THEORY

4-20 mA 11H 0.01% (4mA=0%, 20mA=100%)Gage (pressure/strain)Š120¾ 12H * defined by programmerDisabled channel13H * not applicableBACK COMPATIBILITYIn a

Page 41 - COMMAND SUMMARY

TIMINGThree timing parameters are important from a system integration viewpoint: channel scan rate, communication latency, and host processor speed.

Page 42 - 7409TB (channels 8-15):

byte. This analysis ignores any additional time consumed by the host processor.The read all channels command consumes 740 microseconds maximum as fo

Page 43

BOARD LAYOUTE26E24E22E27E25E23E21E20E18E19J1E17E1E2E3E4E5E6E7E8E9E10E11E12E13E14E15E16P1P2

Page 45 - BACK COMPATIBILITY

INTRODUCTIONFUNCTIONAL DESCRIPTIONThe 7429 interfaces sixteen process sensors directly to the STD bus. Each of the sixteen sensor channels may be ind

Page 46 - COMMUNICATION LATENCY

Measurement time per channel 16.66 msec/channelChannel settling time 5 msecStrain/Pressure gage excitation 10VDC, pulsedResistive sensor excitation 13

Page 47 - PROCESSOR SPEED

SensorsSensor Type Range Resolution Accuracy Noise, 1-sigmaThermocouplesE -270C to 990°C 0.1C 0.2°CJ -210C to 760°C 0.1°C 0.2°CK -270C to 1360°C 0.1°C

Page 48 - BOARD LAYOUT

Strain/Pressure (4-wire bridge) Š 120¾ -100 to +100 mV 5 µV 10 µVDC Voltage Input Range 1 -5 to +5V 200 µV 0.40 mV 0.75 mV Range 2 -500 to +500 mV